Gibralt Spine System Polyaxial Screw Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.

Z-1217-2014

January 06, 2014

Class II

Product summary

Firm
Exactech, Inc.
Event
Event 67486
Status
Terminated
Classification
Class II
Quantity
242
Official record key
device-enforcement:Z-1217-2014

Official wording

Reason: The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.

Code information: size 3.5mm x 10mm, Catalog #05-000-20-3510, Lot Number #35917001. size 3.5mm x 14mm, Catalog #05-000-20-3514, Lot Number 35917002 and 40023002. size 3.5mm x 20mm, Catalog #05-000-20-3520, Lot Number 35917003. size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917004. size 4.0mm x 14mm, Catalog #05-000-20-4014, Lot Number 35917005. size 4.0mm x 20mm, Catalog #05-000-20-4020, Lot Number 35917006. size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917007. 1/4 Shank, size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917008. 1/4 Shank, size 3.5mm x 30mm, Catalog #05-000-20-3530, Lot Number 35917009. 1/4 Shank, size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917010. 1/4 Shank, size 4.0mm x 30mm, Catalog #05-000-20-4030, Lot Number 35917011.

Distribution pattern: USA Distribution including the states of FL, NY, OR, IL, PA, MO and NJ.