Device product

Apollo" Revision/Constrained Knee System: SZ0-16MM AK CCK TIB INSER SZ1-11MM AK CCK TIB INSER SZ1-16MM AK CCK TIB INSER SZ1-22MM AK CCK TIB INSER SZ1-30MM AK CCK TIB INSER SZ2-11MM AK CCK TIB INSER SZ2-13MM AK CCK TIB INSER SZ2-16MM AK CCK TIB INSER SZ2-22MM AK CCK TIB INSER SZ2-30MM AK CCK TIB INSER SZ3-16MM AK CCK TIB INSER SZ3-22MM AK CCK TIB INSER SZ3-25MM AK CCK TIB INSER SZ4-13MM AK CCK TIB INSER SZ4-16MM AK CCK TIB INSER SZ5-11MM AK CCK TIB INSER

Z-1261-2014

February 25, 2014

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 67503
Status
Terminated
Classification
Class II
Quantity
3638 units
Official record key
device-enforcement:Z-1261-2014

Official wording

Reason: A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Code information: Item Number 682000016 682001011 682001016 682001022 682001030 682002011 682002013 682002016 682002022 682002030 682003016 682003022 682003025 682004013 682004016 682005011

Distribution pattern: Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.