Skip to content
Recall Observatory FDA recall evidence

Device product

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Z-1598-2016

April 13, 2016

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 73947
Status
Terminated
Classification
Class II
Quantity
8 systems
Official record key
device-enforcement:Z-1598-2016

Official wording

Reason: The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.

Code information: Model # 10432914 and 10432915 Serial Numbers: 42472, 42489, 42477, 46169, 46178, 46175, 42478, 42485

Distribution pattern: US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.