Skip to content
Recall Observatory FDA recall evidence

Device product

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.

Z-1965-2016

May 10, 2016

Class II

Product summary

Firm
Integra LifeSciences Corp.
Event
Event 74114
Status
Terminated
Classification
Class II
Quantity
71 units
Official record key
device-enforcement:Z-1965-2016

Official wording

Reason: As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Code information: Lot UP15H2, Lot UP09D3, Lot UP07E2

Distribution pattern: US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder