Device product
Ziehm Vision2, Mobile C-arm for Mobile Fluoroscopic Imaging
Z-2644-2016
Product summary
- Event
- Event 74013
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2 devices
- Official record key
device-enforcement:Z-2644-2016
Official wording
Reason: Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
Code information: Serial #'s: 91760 and 91770.
Distribution pattern: US Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.
Derived failure modes
-
Unknown
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.