Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.

Z-1956-2016

May 17, 2016

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 74212
Status
Terminated
Classification
Class II
Quantity
4 units total (1 unit in US)
Official record key
device-enforcement:Z-1956-2016

Official wording

Reason: Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.

Code information: Serial No. 505143, 504805, 501368, 504231

Distribution pattern: US, Morocco, Venezuela

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.