Skip to content
Recall Observatory FDA recall evidence

Device product

Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest On-Site Oral Fluid Drug Screen Device & XALEX One Step Rapid Drug Test. Type of Packaging: 25 individually pouched test devices & collectors in a labeled kit box. The OrALert Oral Fluid Drug Screen Device for AMP/mAMP/COC/OPI/THC/PCP/BZO is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody.

Z-0653-2014

June 03, 2013

Class II

Product summary

Firm
Alere San Diego, Inc.
Event
Event 66870
Status
Terminated
Classification
Class II
Quantity
280,710
Official record key
device-enforcement:Z-0653-2014

Official wording

Reason: Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.

Code information: Lot/Unit Codes: DSD-765-011 (Alere PN 4012350099)  DOA2100709, DOA2110198, DOA2120093, DOA3010031, DOA3010095, DOA3010096, DOA3010393, DOA3010566, DOA3020420. DSD-765-011 (Alere PN 4080350099)  DOA2100672, DOA2120094, DOA2120095, DOA2120158, DOA3010013, DOA3010064, DOA3010097, DOA3010394, DOA3010565, DOA3020557. DSF-765-031 (Alere PN 4584150099)  DOA2100438, DOA2100667, DOA2110073, DOA2110315, DOA2120096, DOA2120738, DOA2120739, DOA3010003, DOA3010004, DOA3010005, DOA3010199, DOA3010202, DOA3020182, DOA3020183. DSF-765-011 (Alere PN 4582290164)  DOA2100250, DOA2110316 , DOA2110317 , DOA2120502, DOA2120503, DOA3010236. DSF-765-011 (Alere PN 4028220211)  DOA2100614. Expiration date, or Expected shelf life: 10/31/14 to 2/28/15. Expected shelf life: 24 months.

Distribution pattern: US Distribution: including states of: AL, CA, PA, and VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.