Device product
VITROS Immunodiagnostic Products HBsAg Controls, Product Code 6800598 For in vitro diagnostic use only. For use in monitoring the performance of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System when used for the in vitro qualitative detection of Hepatitis B Surface Antigen (HBsAg) in human serum and plasma (heparin, EDTA or citrate). The performance of the VITROS Immunodiagnostic Products HBsAg Controls has not been established with any other HBsAg assay.
Z-0313-2017
Product summary
- Event
- Event 74772
- Status
- Terminated
- Classification
- Class III
- Quantity
- 1,621 units
- Official record key
device-enforcement:Z-0313-2017
Official wording
Reason: The signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance.
Code information: Product Code 6800598-Lot #'s: 1) 0860, 2) 0870, 3) 0880, 4) 0890, 5) 0908 Expiration: 1) 29-MAR-17, 2) 12-JUN-17, 3) 02-JUL-17, 4) 23-NOV-17, 5) 29-JAN-18
Distribution pattern: Worldwide Distribution-USA (nationwide) and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.
Derived failure modes
-
Unknown
The signal/cutoff (s/c) results may increase throughout the open-reconstituted timeframe of the allowed 5 days. Increased results may occur on both the negative and positive Controls. The positive Control, although showing elevated results, was still within acceptable performance.