Device product

cobas p 512 Pre-Analytical Instrument calculator/data processing module, for clinical use Product Usage: The cobas 6800/8800 systems support an automated and integrated workflow to run Polymerase Chain Reaction (PCR) based Nucleic Acid Testing (NAT) to be applied in diagnostic laboratories.

Z-0107-2017

July 15, 2016

Class II

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 75186
Status
Terminated
Classification
Class II
Quantity
9014 in total
Official record key
device-enforcement:Z-0107-2017

Official wording

Reason: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Code information: Not applicable

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.