Skip to content
Recall Observatory FDA recall evidence

Device product

Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.

Z-1086-2014

December 18, 2013

Class II

Product summary

Firm
Stryker Spine
Event
Event 67388
Status
Terminated
Classification
Class II
Quantity
21 units
Official record key
device-enforcement:Z-1086-2014

Official wording

Reason: Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.

Code information: D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.