Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Z-2658-2016

July 20, 2016

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 74820
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-2658-2016

Official wording

Reason: After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Code information: 728323 -- Ingenuity Core128 320004 320053 320056

Distribution pattern: US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can