Device product
Ingenuity Core128, Model number 728323 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Z-2658-2016
Product summary
- Event
- Event 74820
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3
- Official record key
device-enforcement:Z-2658-2016
Official wording
Reason: After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
Code information: 728323 -- Ingenuity Core128 320004 320053 320056
Distribution pattern: US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
Derived failure modes
-
Device software or design
software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can