Ingenuity Core, Model number 728321 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Z-2657-2016

July 20, 2016

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 74820
Status
Terminated
Classification
Class II
Quantity
30
Official record key
device-enforcement:Z-2657-2016

Official wording

Reason: After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Code information: 728321 -- Ingenuity Core 310004 310005 310012 310016 310020 310021 310027 310028 310033 310034 310041 310044 310045 310057 310061 310065 310072 310074 310076 310077 310078 310094 310099 310100 310103 310105 310117 310120 310122 310135

Distribution pattern: US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can