GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Z-0840-2015
Product summary
- Event
- Event 69660
- Status
- Terminated
- Classification
- Class II
- Quantity
- 61 units installed in US
- Official record key
device-enforcement:Z-0840-2015
Official wording
Reason: mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Code information: GE OEC Brivo 715 Prime Serial Numbers: B2S1301 6 B2S1301 8 B2S1302 1 B2S13022 B2S13025 B2S13039 B2S13060 B2S13063 B2S13065 B2S14001 B2S14007 B2S14022 B2S14039 B2S1404 0 B2S1404 5 B2S1406 5 B2S1406 6
Distribution pattern: Worldwide Distribution
Derived failure modes
-
Unknown
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).