Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Z-0854-2015
Product summary
- Event
- Event 69478
- Status
- Terminated
- Classification
- Class II
- Quantity
- 44
- Official record key
device-enforcement:Z-0854-2015
Official wording
Reason: The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.
Code information: model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, with serial numbers: 160348 146643 147129 146630 147208 146972 153707 147378 146742 157650 146986 147098 157696 153946 146940 147144 147029 160384 153884 147009 153628 153706 146729 147155 153790 146896 153757 136503 147372 136345 153762 146883 160421 153195 146427 153834 135927 154157 135862 153998 153257 153220 153682 160341
Distribution pattern: USA Distribution to the states of MO, MI, NY, OK, CA, LA, CT, IN, VA, MD, WA, HI, ME, TX, NY, OH, PA and CO.
Derived failure modes
-
Unknown
The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.