Device product
VITROS 3600 Immunodiagnostic System, Catalog Number 6802783, IVD --- Ortho Clinical Diagnostics. For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
Z-0967-2015
Product summary
- Event
- Event 69830
- Status
- Terminated
- Classification
- Class II
- Quantity
- 658 total: USA - 144, Foreign - 514
- Official record key
device-enforcement:Z-0967-2015
Official wording
Reason: Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Code information: Software Version 3.1 & Below; Serial Numbers J36000101 - J36000834 (the serial numbers are sequential; J Numbers are analogous to serial numbers)
Distribution pattern: Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Derived failure modes
-
Device software or design
Software Anomaly: the firm has identified an anomaly