Device product
DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715
Z-0092-2015
Product summary
- Firm
- Synthes, Inc.
- Event
- Event 69432
- Status
- Terminated
- Classification
- Class II
- Quantity
- 11
- Official record key
device-enforcement:Z-0092-2015
Official wording
Reason: Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a patient is exposed to latex particles from the adhesive tape, or a systemic reaction such as anaphylaxis in patients with latex allergy.
Code information: Radiolucent Retractor for Upper Cervical Spine part number 387.580, lot number 3620715.
Distribution pattern: US Distribution including the states of PA, NC, TN, NY, WI, WA and MI.
Derived failure modes
-
Unknown
Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a patient is exposed to latex particles from the adhesive tape, or a systemic reaction such as anaphylaxis in patients with latex allergy.