Skip to content
Recall Observatory FDA recall evidence

Device product

Target Detachable Coils: TARGET HELICAL NANO 3MM X 6CM MODEL Number:M0035453060 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Z-2186-2016

June 06, 2016

Class II

Product summary

Firm
Stryker Neurovascular
Event
Event 74425
Status
Terminated
Classification
Class II
Quantity
45,302 (8,188 in the U.S.) - total, all model numbers
Official record key
device-enforcement:Z-2186-2016

Official wording

Reason: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Code information: MODEL Number:M0035453060; Lot Number/Expiration Date: 18499486 31-Jul-18 18603584 30-Sep-18 19086497 31-May-19.

Distribution pattern: Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.