Skip to content
Recall Observatory FDA recall evidence

Device product

Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.

Z-0171-2020

August 01, 2018

Class II

Product summary

Firm
ICU Medical Inc
Event
Event 83480
Status
Terminated
Classification
Class II
Quantity
47,152 devices
Official record key
device-enforcement:Z-0171-2020

Official wording

Reason: A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.

Code information: All Plum 360 Infusers with software versions 15.10.00.010/.011/.012 and 15.11.00.017 when used with an ICU Medical MedNet CDL

Distribution pattern: US nationwide distribution, and Canada

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software malfunction