Device product
CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS
Z-1696-2019
Product summary
- Event
- Event 82664
- Status
- Terminated
- Classification
- Class II
- Quantity
- 68 units
- Official record key
device-enforcement:Z-1696-2019
Official wording
Reason: There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Code information: a) 0660000: CyberKnife Treatment Delivery System Serial Numbers: C0434, C0436, C0437, C0438, C0440, C0441, C0443, C0444, C0445, C0446, C0447, C0448, C0449, C0450, C0453, C0454, C0456, C0457, C0458, C0460, C0464, C0465, C0466, C0472, C0475 b) 053101-010: CyberKnife M6 FI Treatment Delivery System Serial Numbers: C0334, C0381 c) 053201-010: CyberKnife M6 FM Treatment Delivery System Serial Numbers: C0363 d) 053301-010: CyberKnife M6 FIM Treatment Delivery System Serial Numbers: C0337, C0343, C0349, C0352, C0358, C0370, C0371, C0378, C0384, C0387, C0389, C0390, C0391, C0393, C0409, C0424, C0426, C0427, C0428, C0433, C0439 e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System Serial Numbers: C0330, C0332, C0342, C0344, C0345, C0348, C0359, C0360, C0364, C0367, C0372, C0394, C0398, C0400, C0402, C0404, C0415, C0419 f) 053501-010: CyberKnife Matrix Tier F TDS Serial Numbers: C0396
Distribution pattern: US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
Derived failure modes
-
Unknown
There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.