Device product

Target Detachable Coils: TARGET HELICAL NANO 1.5 MM X 1 CM MODEL Number:M0035431510 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Z-2168-2016

June 06, 2016

Class II

Product summary

Firm
Stryker Neurovascular
Event
Event 74425
Status
Terminated
Classification
Class II
Quantity
45,302 (8,188 in the U.S.) - total, all model numbers
Official record key
device-enforcement:Z-2168-2016

Official wording

Reason: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Code information: MODEL Number: M0035431510; Lot Number/Expiration Date: 18581295 30-Sep-18 18800256 31-Jan-19 18775373 31-Dec-18 18744204 30-Nov-18 18812788 31-Jan-19 18607578 30-Sep-18 18602655 30-Sep-18 19003141 30-Apr-19 18935435 31-Mar-19 18923481 31-Mar-19 18953907 31-Mar-19 18883412 28-Feb-19 18852906 28-Feb-19 19056331 31-May-19.

Distribution pattern: Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.