Device product
10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Z-0493-2018
Product summary
- Event
- Event 78692
- Status
- Terminated
- Classification
- Class II
- Quantity
- 10 bags, 100 units each
- Official record key
device-enforcement:Z-0493-2018
Official wording
Reason: The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Code information: DATE - MVT 25.08.2017
Distribution pattern: Nationwide
Derived failure modes
-
Unknown
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.