Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide
Z-1538-2012
Product summary
- Event
- Event 61830
- Status
- Terminated
- Classification
- Class II
- Quantity
- 11 - US
- Official record key
device-enforcement:Z-1538-2012
Official wording
Reason: When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie
Code information: Unavailable
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie