Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other devices via Flexport interfaces. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Spacelabs Healthcare Xprezzon Bedside Monitor may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitor allows network based applications to open windows and display information the Xprezzon and other networked monitors. The Spacelabs Healthcare Xprezzon Bedside Monitor is designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Xprezzon Bedside Monitors is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

Z-2019-2012

June 12, 2012

Class II

Product summary

Firm
Spacelabs Healthcare, Llc
Event
Event 62437
Status
Terminated
Classification
Class II
Quantity
21 units
Official record key
device-enforcement:Z-2019-2012

Official wording

Reason: There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.

Code information: 040-1632-00; 1393-000013; 1393-000039; 1393-000040; 1393-000041; 1393-000341; 1393-000382; 1393-000383; 1393-000384; 1393-000385; 1393-000386; 1393-000387; 1393-000388; 1393-000808; 1393-000831; 1393-001374; 1393-001375; 1393-001453; 1393-001990; 1393-001991; and 1393-001992.

Distribution pattern: Worldwide Distribution - USA including Florida, Iowa, Oklahoma, South Caroline and the countries of India, Germany and Panama.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.