Device product
CervAlign Anterior Cervical Plate System-Cervalign, 3Level, 57mm Material Number: 66-357 - Product Usage: intended for anterior cervical fixation (C2-C7) for the following conditions: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudathrosis, and failed previous fusion).
Z-1225-2021
Product summary
- Event
- Event 87251
- Status
- Ongoing
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-1225-2021
Official wording
Reason: Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision
Code information: Batch Number: 333152 333152 338845 341493 347126 351531 355537 359848 364534 364971 369364 370407 378060 380152
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision