Device product

VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997

Z-0061-2019

July 12, 2018

Class II

Product summary

Firm
Ortho Clinical Diagnostics Inc
Event
Event 80717
Status
Ongoing
Classification
Class II
Quantity
68230
Official record key
device-enforcement:Z-0061-2019

Official wording

Reason: The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

Code information: 5535 5540 5550 5560 5570 5580 5590 5600 5610 5615 5620 5630 5640 5650 5655 5660 5665 5670 5675 5678 5680 5685 5690

Distribution pattern: Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.