Device product

Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations: Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS The JANUS Automated Workstation is an automated, programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows. The NGS Express is an automated, programmable liquid handling instrument suitable for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation workflows

Z-0747-2017

June 24, 2016

Class II

Product summary

Firm
PerkinElmer, Inc
Event
Event 75508
Status
Terminated
Classification
Class II
Quantity
595 trays (571,200 tips)
Official record key
device-enforcement:Z-0747-2017

Official wording

Reason: Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing

Code information: Lot Numbers:15470394, 15471775, 1601067, 16071147

Distribution pattern: Worldwide Distribution - US including CA, MA, NY, IA, OR , PA, WA and Internationally to Belgium, Canada, Finland, France, Germany, Japan, Netherlands, Singapore, and UK

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing