ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay.

Z-1260-2020

November 19, 2019

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 84413
Status
Terminated
Classification
Class II
Quantity
844
Official record key
device-enforcement:Z-1260-2020

Official wording

Reason: Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.

Code information: Siemens Material Number (SMN): ADVIA Chemistry: 10335897 REF: 05006455 Lot# 453610 Expiration Date: 2020-01 UDI Number: ADVIA: (01)00630414526843(10)453610(17)20200128

Distribution pattern: Worldwide distribution  US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WY and countries of UK, CA, PH, MX, VN, TW, TH, IN, CO, AE, AR, AU, BD, BE, BG, BR, CL, CZ, DE, DK, EG, ES, FI, FR, GR, HU, IE, IT, KR, MY, NL, PL, PT, RO, RU, SE, SG.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.