Device product
Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin, troponin I and B-type natriuretic peptide in EDTA anticoagulated whole blood and plasma specimens.
Z-1928-2012
Product summary
- Event
- Event 61927
- Status
- Terminated
- Classification
- Class I
- Quantity
- 16,962 kits
- Official record key
device-enforcement:Z-1928-2012
Official wording
Reason: Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Code information: Phase 1: W49604B, W49606B, W49607B, W49611B, W49612B, W49613B, W49614B, W49615B, W49616B, W49617B, W49620B, W49622B, W49623B, W49624B, W49625B, W49627B, W49628B, W49629B, W49630B, W49631B, W49633B, W49634B, W49635B, W49638B, W49639B, W50726B, W50727B, W50728B, W50731B, W50733B, W50734B, W50738B, and W50739B. Phase 2: W49609B, W49610B, W49618B, W49619B, W49626B, W49632B, W49640B, W50737B, and W50744B.
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.