Device product
INFINITY DUAL HEMO MCable Pod
Z-0942-2018
Product summary
- Event
- Event 79209
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2,760
- Official record key
device-enforcement:Z-0942-2018
Official wording
Reason: Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Code information: Have the affected revision indexes: 15 or 16 (found on the product label)
Distribution pattern: Nationally
Derived failure modes
-
Unknown
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.