Siemens Dimension(R) Tacrolimus Flex(R) reagent cartridge (DF107). The TACR method is used as an in vitro diagnostic test.

Z-0957-2013

November 27, 2012

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 63767
Status
Terminated
Classification
Class II
Quantity
2,976
Official record key
device-enforcement:Z-0957-2013

Official wording

Reason: Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range. Individual patients, particularly samples with lower TACR values

Code information: Lot numbers FA3085, exp 2013-03-26; and FA3316, exp 2013-11-12

Distribution pattern: Product was distributed within the United States to AL, AZ, CA, DC, FL, LA, MD, MI, MO, NJ, NM, NC, OH, OR, PA, TN, TX, UT, VA, and WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Siemens has confirmed customer complaints of low patient sample recovery with Dimension(R) TACR Flex(R) lot FA3085. Quality Control (QC) materials have not exhibited low recovery of the same magnitude. Internal method regression testing versus LC/MS has shown an average 25% low bias compared to LC/MS across the assay range. Individual patients, particularly samples with lower TACR values