Device product
DC power cable (p/n 3010556) supplied with the MEDRAD(R) Veris MR Vital Signs Monitor: Product Usage: Monitor, Physiological, Patient
Z-1586-2013
Product summary
- Firm
- Medrad Inc
- Event
- Event 65411
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1,066
- Official record key
device-enforcement:Z-1586-2013
Official wording
Reason: The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.
Code information: Catalog numbers 3011992, 3011992W, 3011993, 3011993W, 3011994, 3011994W, 3011996, 3011996W, 3014268 and 3014268W.
Distribution pattern: Worldwide Distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and the countries of: AE, AU, AZ, BN, BR, CA, CH, CL, CO, DE, DK, DO, DZ, EG, ES, FR, GB, HK, HN, HR, HU, IN, IT, JO, JP, KR, LB, MU, MX, MY, NL, PH, PS, SA, SG, TH, TR, TW, UY, VE, and ZA.
Derived failure modes
-
Unknown
The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.