Skip to content
Recall Observatory FDA recall evidence

Device product

Dimension Vista Systems LOCI Thyroid Stimulating Hormone; the TSH method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone in human serum and plasma on the Dimension Vista System.

Z-1708-2015

April 20, 2015

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 71096
Status
Terminated
Classification
Class II
Quantity
2256
Official record key
device-enforcement:Z-1708-2015

Official wording

Reason: Dimension Vista LOCI Thyroid Stimulating Hormone (TSH) Lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a Flex well.

Code information: Catalog Number K6412, Material Number 10445104, Lot Number 14237AA, exp. 2015-08-25.

Distribution pattern: Worldwide Distribution - Nationwide and to the countries of : Australia, Japan, New Zealand, Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, Great Britain, Italy, Netherlands, Portugal, and Slovakia.