Skip to content
Recall Observatory FDA recall evidence

Device product

Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017

Z-0625-2018

June 08, 2017

Class II

Product summary

Firm
Roche Diabetes Care, Inc.
Event
Event 79001
Status
Terminated
Classification
Class II
Quantity
67,040 unique users of version 2.1.0
Official record key
device-enforcement:Z-0625-2018

Official wording

Reason: Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.

Code information: Software Versions: 1.2.0, 1.2.2, 1.2.3, 2.0.0, 2.0.1, and 2.1.0

Distribution pattern: US Nationwide

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software bug