Device product
Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000 Control Unit and Energy Transfer Pads is a thermal regulating system, indicated for monitoring and controlling patient temperature.
Z-2181-2012
Product summary
- Event
- Event 62596
- Status
- Terminated
- Classification
- Class II
- Quantity
- 129 units
- Official record key
device-enforcement:Z-2181-2012
Official wording
Reason: A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.
Code information: Catalogue #01000-01 and 01000-01R, Units with 4-Digit Serial Numbers: Lot #4297, 4324, 4332, 4519, 4559, 4572; Units with 10 Character Serial Numbers: SA08017027, SA08017028, SA08017029, SA08017030, SA08017031, SA08017032, SA08017033, SA08017034, SA08017035, SA08017036, SA08017037, SA08017038, SA08017039, SA08017040, SA08017041, SA08017042, SA08017043, SA08017044, SA08017045, SA08017046, SA08017047, SA08017048, SA08017049, SA08343100, SA11137004, SA11137005, SA11137006, SA11137007, SA11137008, SA08094124, SA08094125, SA08094126, SA08094127, SA08094128, SA08094129, SA08094130, SA08094131, SA08094132, SA08094133, SA08094134, SA08094135, SA08094136, SA08094137, SA08094138, SA08094139, SA08094140, SA08094141, SA08094142, SA08094143, SA08094144, SA08094145, SA08094146, SA08094147, SA08094148, SA08094149, SA08094150, SA08094151, SA08094152, SA08094153, SA09020004, SA09020005, SA09020007, SA09020008, SA09020010, SA09020013, SA11298011, SA11298012, SA11298013, SA11298014, SA11298015, SA11298016, SA11298017, SA11298018, SA11298019, SA08136001, SA08136002, SA08136003, SA08136004, SA08136005, SA08136006, SA08136007, SA08136008, SA08136009, SA08136010, SA09043001, SA09043002, SA09043003, SA09043004, SA09043007, SA09043008, SA09043009, SA09043010, SA09043010, SA09043011, SA09043012, SA09043013, SA09043014, SA09043015, SA09043016, SA09043017, SA09043018, SA09043019, SA09043020, SA08343061, SA08343082, SA08343083, SA08343084, SA08343085, SA08343086, SA08343087, SA08343088, SA08343089, SA08343090, SA08343091, SA08343092, SA08343093, SA08343094, SA08343095, SA08343096, SA10337006, SA10337007, SA10337008, SA10337009 and SA10337010.
Distribution pattern: Nationwide Distribution-including the states of AZ, CA, CO, DE, FL, GA, IA, ID, IN, IL, KS, KY, LA, MA, MD, ME, MO, MS, NE, NM, OH, OR, PA, RI, SC, TN, TX, VA, WI and WV.
Derived failure modes
-
Unknown
A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.