Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured by Titan Spine, LLC. Endo TCS driver is one of the instruments within the surgical tray for use with the Interbody Fusion Device(IBD).

Z-1324-2015

February 20, 2015

Class II

Product summary

Firm
TITAN SPINE, LLC
Event
Event 70576
Status
Terminated
Classification
Class II
Quantity
26
Official record key
device-enforcement:Z-1324-2015

Official wording

Reason: The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.

Code information: G150101

Distribution pattern: US Distribution to the states of : AZ, CA, CT, FL, GA, ID, IL, KY, MD, MO, NV, NY, OH, OK, OR, TX, UT and WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.