Merge Cardio software. Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.

Z-0729-2017

March 28, 2016

Class II

Product summary

Firm
Merge Healthcare, Inc.
Event
Event 75825
Status
Terminated
Classification
Class II
Quantity
198 sites potentially have the effected versions
Official record key
device-enforcement:Z-0729-2017

Official wording

Reason: If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession number (but still associated to the correct patient).

Code information: Versions: 8.30.0, 8.30.1, 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 9.0.9, 10.0, 10.0.1, 10.1, and 10.1.1

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession number (but still associated to the correct patient).