Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Z-2383-2012
Product summary
- Event
- Event 63033
- Status
- Terminated
- Classification
- Class II
- Quantity
- 221 Units
- Official record key
device-enforcement:Z-2383-2012
Official wording
Reason: Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
Code information: LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H294635 ,H294695 ,H294758 ,H140891 ,H294636 ,H294696 ,H294759 ,H140892 ,H294637 ,H294697 ,H294760 ,H140893 ,H294638 ,H294698 ,H294761 ,H140895 ,H294639 ,H294699 ,H294762 ,H140896 ,H294640 ,H294700 ,H294763 ,H140897 ,H294641 ,H294701 ,H294764 ,H140898 ,H294642 ,H294702 ,H294765 ,H140899 ,H294643 ,H294704 ,H294766 ,H140900 ,H294644 ,H294705 ,H294767 ,H181175 ,H294645 ,H294706 ,H294768 ,H181176 ,H294646 ,H294707 ,H294769 ,H181177 ,H294647 ,H294708 ,H294770 ,H181178 ,H294648 ,H294709 ,H294771 ,H181179 ,H294649 ,H294710 ,H294772 ,H181180 ,H294650 ,H294711 ,H181181 ,H294651 ,H294712 ,H181182 ,H294652 ,H294713 ,H181183 ,H294653 ,H294714 ,H181184 ,H294654 ,H294715 ,H181185 ,H294655 ,H294716 ,H181186 ,H294656 ,H294717 ,H181187 ,H294657 ,H294718 ,H181188 ,H294658 ,H294719 ,H181189 ,H294659 ,H294720 ,H181190 ,H294660 ,H294722 ,H181191 ,H294661 ,H294723 ,H181192 ,H294662 ,H294724 ,H182000 ,H294663 ,H294726 ,H182001 ,H294664 ,H294727 ,H182002 ,H294665 ,H294728 ,H182003 ,H294666 ,H294729 ,H182005 ,H294667 ,H294730 ,H294607 ,H294668 ,H294731
Distribution pattern: Worldwide Distribution.
Derived failure modes
-
Unknown
Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.