Device product
Precise Digital Accelerator Delivery of radiation to defined target volumes
Z-2405-2012
Product summary
- Firm
- Elekta, Inc.
- Event
- Event 62917
- Status
- Terminated
- Classification
- Class II
- Quantity
- 110
- Official record key
device-enforcement:Z-2405-2012
Official wording
Reason: There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
Code information: 151587, 105932, 151574, 151363, 151076, 151110, 151369, 151504, 105941, 151418, 151674, 151566, 151259, 151636, 151568, 151390, 151258, 151584, 151391, 151317, 151522, 151055, 151423, 151507, 151623, 151627, 151640, 151479, 151340, 151178, 151467, 151514, 151051, 151411, 151216, 151430, 151173, 151616, 151595, 151141, 151433, 151611, 151151, 151653, 151515, 151360, 151675, 151538, 151024, 151523, 151521, 151260, 151519, 151432, 151410, 151371, 151084, 151435, 151387, 151530, 151328, 151176, 151080, 151626, 151567, 151598, 151601, 151269, 151396, 151557, 151558, 105982, 151160, 151634, 151394, 151398, 151540, 151632, 151446, 151301, 151038, 151441, 151256, 151168, 151156, 151157, 151155, 151579, 151412, 105984, 151053, 151359, 151585, 151326, 151327, 151443, 151250, 151130, 151245, 151416, 151664, 151672, 151074, 151352, 151298, 151545, 151628, 105803, 151362, 151073
Distribution pattern: Nationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IA, MD, MA, MI, MN, NE, NV, NJ, NM, NY, ND, OH, OK, OR, PA, VA, SC, TN TX, WA, WV, WI and Puerto
Derived failure modes
-
Unknown
There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.