Device product

Natrelle133 Plus Tissue Expander With/Without Suture Tabs, BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style numbers: STYLE 133P-FV: 133P-FV-11, 133P-FV-12, 133P-FV-13, 133P-FV-14, 133P-FV-15, 133P-FV-16; STYLE 133P-MV: 133P-MV-11, 133P-MV-12, 133P-MV-13, 133P-MV-14, 133P-MV-15, 133P-MV-16; STYLE 133P-LV: 133P-LV-11, 133P-LV-12, 133P-LV-13, 133P-LV-14, 133P-LV-15, 133P-LV-16; STYLE 133P-FX: 133P-FX-11, 133P-FX-12, 133P-FX-13, 133P-FX-14, 133P-FX-15, 133P-FX-16; STYLE 133P-MX:133P-MX-11,133P-MX-12, 133P-MX-13, 133P-MX-14, 133P-MX-15, 133P-MX-16; STYLE 133P-SX: 133P-SX-11, 133P-SX-12, 133P-SX-13, 133P-SX-14, 133P-SX-15, 133P-SX-16; STYLE 133P-SV: 133P-SV-11,133P-SV-12, 133P-SV-13, 133P-SV-14, 133P-SV-15, 133P-SV-16; STYLE 133P-FV-T:133P-FV-11-T, 133P-FV-12-T, 133P-FV-13-T, 133P-FV-14-T, 133P-FV-15-T, 133P-FV-16-T; STYLE 133P-MV-T: 133P-MV-11-T, 133P-MV-12-T, 133P-MV-13-T, 133P-MV-14-T, 133P-MV-15-T, 133P-MV-16-T; STYLE 133P-LV-T: 133P-LV-11-T, 133P-LV-12-T, 133P-LV-13-T, 133P-LV-14-T, 133P-LV-15-T, 133P-LV-16-T; STYLE 133P-FX-T: 133P-FX-11-T, 133P-FX-12-T, 133P-FX-13-T, 133P-FX-14-T, 133P-FX-15-T, 133P-FX-16-T; STYLE 133P-MX-T: 133P-MX-11-T, 133P-MX-12-T, 133P-MX-13-T, 133P-MX-14-T, 133P-MX-15-T, 133P-MX-16-T; STYLE 133P-SX-T: 133P-SX-11-T, 133P-SX-12-T, 133P-SX-13-T, 133P-SX-14-T, 133P-SX-15-T, 133P-SX-16-T; STYLE 133P-SV-T: 133P-SV-11-T, 133P-SV-12-T, 133P-SV-13-T, 133P-SV-14-T, 133P-SV-15-T, 133P-SV-16-T These products may have been sold under the McGhan, Inamed or NATRELLE brands. Product Usage: intended for temporary subcutaneous implantation and require periodic, incremental inflation with sterile saline for injection until the desired amount of tissue is developed

Z-2456-2019

July 24, 2019

Class I

Product summary

Firm
Allergan PLC
Event
Event 83500
Status
Ongoing
Classification
Class I
Quantity
4,026,287 Breast Implants and Tissue Expanders Combined in total
Official record key
device-enforcement:Z-2456-2019

Official wording

Reason: The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.

Code information: ALL LOTS; Product Code/UDI STYLE 133P-FV:133P-FV-11 10888628009646, 133P-FV-11-T 10888628009653, 133P-FV-12 10888628009660, 133P-FV-12-T 10888628009677, 133P-FV-13 10888628009684, 133P-FV-13-T 10888628009691, 133P-FV-14 10888628009707, 133P-FV-14-T 10888628009714, 133P-FV-15 10888628009721, 133P-FV-15-T 10888628009738, 133P-FV-16 10888628009745; STYLE 133P-MV: 133P-MV-11 10888628010000, 133P-MV-12 10888628010024, 133P-MV-13 10888628010048, 133P-MV-14 10888628010062, 133P-MV-15 10888628010086, 133P-MV-16 10888628010109; STYLE 133P-LV: 133P-LV-11 10888628009882, 133P-LV-12 10888628009905, 133P-LV-13 10888628009929, 133P-LV-14 10888628009943, 133P-LV-15 10888628009967, 133P-LV-16 10888628009981; STYLE 133P-FX: 133P-FX-11 10888628009769, 133P-FX-12 10888628009783, 133P-FX-13 10888628009806, 133P-FX-14 10888628009820, 133P-FX-15 10888628009844, 133P-FX-16 10888628009868; STYLE 133P-MX: 133P-MX-11 10888628010123, 133P-MX-12 10888628010147, 133P-MX-13 10888628010161, 133P-MX-14 10888628010185, 133P-MX-15 10888628010208, 133P-MX-16 10888628010222; STYLE 133P-SX: 133P-SX-11 10888628010369, 133P-SX-12 10888628010383, 133P-SX-13 10888628010406, 133P-SX-14 10888628010420, 133P-SX-15 10888628010444, 133P-SX-16 10888628010468; STYLE 133P-SV: 133P-SV-11 10888628010246, 133P-SV-12 10888628010260, 133P-SV-13 10888628010284, 133P-SV-14 10888628010307, 133P-SV-15 10888628010321, 133P-SV-16 10888628010345; STYLE 133P-FV-T: 133P-FV-11-T 10888628009653, 133P-FV-12-T 10888628009677, 133P-FV-13-T 10888628009691, 133P-FV-14-T 10888628009714, 133P-FV-15-T 10888628009738, 133P-FV-16-T 10888628009752; STYLE 133P-MV-T: 133P-MV-11-T 10888628010017, 133P-MV-12-T 10888628010031, 133P-MV-13-T 10888628010055, 133P-MV-14-T 10888628010079, 133P-MV-15-T 10888628010093, 133P-MV-16-T 10888628010116; STYLE 133P-LV-T: 133P-LV-11-T 10888628009899, 133P-LV-12-T 10888628009912, 133P-LV-13-T 10888628009936, 133P-LV-14-T 10888628009950, 133P-LV-15-T 10888628009974, 133P-LV-16-T 10888628009998; STYLE 133P-FX-T: 133P-FX-11-T 10888628009776, 133P-FX-12-T 10888628009790, 133P-FX-13-T 10888628009813, 133P-FX-14-T 10888628009837, 133P-FX-15-T 10888628009851, 133P-FX-16-T 10888628009875; STYLE 133P-MX-T: 133P-MX-11-T 10888628010130, 133P-MX-12-T 10888628010154, 133P-MX-13-T 10888628010178, 133P-MX-14-T 10888628010192, 133P-MX-15-T 10888628010215, 133P-MX-16-T 10888628010239; STYLE 133P-SX-T: 133P-SX-11-T 10888628010376, 133P-SX-12-T 10888628010390, 133P-SX-13-T 10888628010413, 133P-SX-14-T 10888628010437, 133P-SX-15-T 10888628010451, 133P-SX-16-T 10888628010475; STYLE 133P-SV-T: 133P-SV-11-T 10888628010253, 133P-SV-12-T 10888628010277, 133P-SV-13-T 10888628010291, 133P-SV-14-T 10888628010314, 133P-SV-15-T 10888628010338, 133P-SV-16-T 10888628010352

Distribution pattern: Worldwide Distribution and US Nationwide All fifty (50) US States, the US Virgin Islands and Puerto Rico. (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY). OUS: Bermuda, Canada, Chile, Japan, Korea, Thailand, Taiwan and Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.