Device product
Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications
Z-2384-2012
Product summary
- Event
- Event 62884
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4 systems
- Official record key
device-enforcement:Z-2384-2012
Official wording
Reason: Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
Code information: Lot or Serial Numbers With affected M-cabinet CXA; Serial number; 11-K0003; 12-B0001, 12-B0002; 12-B0003
Distribution pattern: Worldwide Distribution - USA including OH and WA and the country of Canada
Derived failure modes
-
Unknown
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.