Skip to content
Recall Observatory FDA recall evidence

Device product

BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamlined and automated. Routine tasks such as daily cleaning and shutdown can be programmed to occur automatically. The BD FACSuite clinical software allows users to run BD IVD assays.

Z-0792-2019

January 02, 2019

Class III

Product summary

Firm
Becton, Dickinson and Company, BD Biosciences
Event
Event 81915
Status
Terminated
Classification
Class III
Quantity
3 devices
Official record key
device-enforcement:Z-0792-2019

Official wording

Reason: A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.

Code information: UDI: (01)00382906628787(11)180508(21)Z662878000056 serial number: Z662878000056 Work Station SN 2UA80820ZZ UDI: (01)00382906628787(11)180301(21)Z662878000029 serial number: Z662878000029 Work Station SN 2UA80820ZW UDI: (01)00382903389605(11)180315(21)11662382000007 serial number: R662382000007 Work Station SN 2UA8082101

Distribution pattern: US Distribution in MD.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.