Device product
BioFlo PICC with ENDEXO Technology and PASV Valve Technology Maximal Barrier Nursing Kit, 5F-55 cm, UPN Product No. H965750191, REF/Catalog No. 75-019, STERILE --- Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood and for power injection of contrast media.
Z-0026-2015
Product summary
- Event
- Event 69085
- Status
- Terminated
- Classification
- Class II
- Quantity
- 300 units/pouches
- Official record key
device-enforcement:Z-0026-2015
Official wording
Reason: NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.
Code information: Batch/Lot Nos. 4745699 (exp. 31-Jan-16) and 4760237 (31-Jan-16)
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.