Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens syngo Imaging XS, version VA70A or higher Syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo Imaging XS also supports storage and archiving of Structured DICOM reports. Syngo Imaging XS optionally uses a variety of advanced post processing applications.

Z-2111-2014

June 30, 2014

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 68769
Status
Terminated
Classification
Class II
Quantity
141
Official record key
device-enforcement:Z-2111-2014

Official wording

Reason: Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d

Code information: model number: 10496279

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel Lens, Edge enhancement and Histograms. This potential malfunction may occur for images with a pixel depth greater than 12 bit. All values calculated for images with a pixel d