Device product
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Z-2835-2018
Product summary
- Event
- Event 80691
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1
- Official record key
device-enforcement:Z-2835-2018
Official wording
Reason: An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Code information: UDI Number: 00886874114193. Serial Number: SK1703.
Distribution pattern: U.S.
Derived failure modes
-
Unknown
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.