Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.
Z-0149-2015
Product summary
- Firm
- Tecan US, Inc.
- Event
- Event 69369
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-0149-2015
Official wording
Reason: Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.
Code information: Part Number: 10641100, SN: 1407006311.
Distribution pattern: US Distribution to IL.
Derived failure modes
-
Unknown
Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.