TRINIAS for diagnostic imaging and interventional procedures in cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.

Z-0864-2019

January 14, 2019

Class II

Product summary

Firm
Shimadzu Medical Systems Usa Com
Event
Event 82002
Status
Terminated
Classification
Class II
Quantity
13 units
Official record key
device-enforcement:Z-0864-2019

Official wording

Reason: Two issues: Event 1: Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure. Event 2: In rare cases a synchronization error may occur in the data transmission circuit in the device due to external noise. The visibility of the displayed image may be corrupted (images become bit-shifted abnormal images). If this event occurs, it may become difficult to see the object, which may hinder the examination and treatment.

Code information: Serial Numbers: 41B247746001, 41B24774C001, 41E58C757001, 41E58C765001, 41E58C766001, 41E58C769001, 41E58C76B001, 41E58C773001, 41E58C775001, 41E58C776001, 41E58C777001, 41E58C977001, 41EFCD77C001

Distribution pattern: FL, MT, CT, OH, IL, TX, MS, LA, SC

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two issues: Event 1: Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure. Event 2: In rare cases a synchronization error may occur in the data transmission circuit in the device due to external noise. The visibility of the displayed image may be corrupted (images become bit-shifted abnormal images). If this event occurs, it may become difficult to see the object, which may hinder the examination and treatment.