Skip to content
Recall Observatory FDA recall evidence

Device product

General Cosmetic Pack FCSC, Code 900-3113, contains: (1) COVER MAYO STAND REINFORCED LIF (10) LAP SPONGE PREWASH 18" X 18" XRD LIF (2) GOWN LARGE STANDARD SMS AAMIIII (1) TABLE COVER REINFORCED 50" X 90" LIF (2) TOWEL ABSORBENT 15" X 20" LIF (1) YANKAUER SUCTION TUBE WITHOUT VENT LIF (1) SYRINGE BULB 60CC LIF (1) BLADE SURGICAL #15 CARBON STEEL (1) NEEDLE HYPODERMIC 18G X 1% LIF (6) TOWEL CLOTH HUCK BLUE LIF (4) DRAPE UTILITY WITH TAPE L/F (1) NEEDLE HYPODERMIC 25 X 1% L/F (1) SYRINGE 20CC WITHOUT NEEDLE LUER LOCK LIF (2) SYRINGE 10CC WITHOUT NNEDEL LUER LOCK LIF (2) NEEDLE SPINAL ANESHT 20G X 3%" LIF (1) SKIN MARKER WITH RULER LIF (1) COUNTER NEEDLE & BLADE 20C FOAM/MAG L/F (1) LITE GLOVE (1) DRAPE BREAST/CHEST SMS LIF (1) TUBE SUCTION CONNECT%" X 12' LIF (10) GAUZE SPONGE 4" X 4" 16PLY L/F (2) GLOVE MEDIUM FREETOOUCH VYNIL POWDER FREE (1) TRAY 3/COMPARTMENT (1) WRAPPER 24" X 24" L/F (2) PK. SURG. GLOVE# 7% ENCORE POWDER FREE LATEX Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Z-2523-2014

May 20, 2014

Class I

Product summary

Firm
Customed, Inc
Event
Event 68536
Status
Terminated
Classification
Class I
Quantity
2 lots/ 18 units, multiple units per lot
Official record key
device-enforcement:Z-2523-2014

Official wording

Reason: Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Code information: 900-3113, 2 lots: 131110435 131210803

Distribution pattern: Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    packaging integrity may be compromised
  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    loss of sterility