Device product
MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system.
Z-2386-2012
Product summary
- Event
- Event 63014
- Status
- Terminated
- Classification
- Class II
- Quantity
- 328 units
- Official record key
device-enforcement:Z-2386-2012
Official wording
Reason: On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
Code information: System Code #, 728131, Serial #: EP16E090001-EP16E090004, EP16E090006-EP16E090015, EP16E100001- EP16E100004, EP16E100006, EP16E100008- EP16E100063, EP16E100067, EP16E100069- EP16E100113, EP16E110001- EP16E110065, EP16E110067-EP16E110070, EP16E110074-EP16E110075, EP16E110077-EP16E110078, EP16E110080, EP16E110083- EP16E110085, EP16E110088- EP16E110112, EP16E110115- EP16E110119, EP16E120001- EP16E120014, EP16E120016, EP16EB110003- EP16ER100007, EP16ER110003, EP16ER110006, P16C080001, P16C080002, P16C090001, P16C100001-P16C100015, P16C100017, P16C100018, P16C110001- P16C110008, P16C110011-P16C110037, P16C120001-P16C120003, P16C120006, P16C120007, P16E080005, P16E080007, P16E080008, P16E080012, P16E090002-P16E090021, P16E090024-P16E090028, P16E100002, P16EF80010, P16EM08003, and P16EM09002-P16EM09004. System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003-EP16ER100005, EP16ER100008, EP16ER110004, EP16ER110005, and EP16ER110007-EP16ER110024.
Distribution pattern: Worldwide distribution, Nationwide (US) including the states AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, and VA and the countries of Albania, Argentina, Australia, Belgium, Brazil, Chile, China, Colombia, Dominican Republic, Ecuador, France, Germany, Greece, India, Indonesia, Ireland, Italy, Jordan, Lebanon, Malawi, Mexico, Nepal, Netherlands, Oman, Palestine, Philippines, Poland, Romania, Russia, Saudi Arabia, South Africa, South Korea, Spain, Switzerland, Turkey, Uganda, and Venezuela.
Derived failure modes
-
Unknown
On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.