Device product
CE17TKFS EPIDURAL TRAY, Material Number 332098 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.
Z-3344-2018
Product summary
- Event
- Event 80811
- Status
- Terminated
- Classification
- Class II
- Quantity
- 24810
- Official record key
device-enforcement:Z-3344-2018
Official wording
Reason: The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Code information: Batch: 61586632, 61591151, 61594892, 61599653, 61604114, 61608068, 61612419, 61615748, 61619582
Distribution pattern: Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Guatemala and Kuwait.
Derived failure modes
-
Unknown
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.