Device product
CE18T CONT EPIDURAL SET, Material Number 332200 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.
Z-3350-2018
Product summary
- Event
- Event 80811
- Status
- Terminated
- Classification
- Class II
- Quantity
- 10500
- Official record key
device-enforcement:Z-3350-2018
Official wording
Reason: The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Code information: Batch: 61599578, 61602043, 61605370, 61610383, 61613469, 61615920
Distribution pattern: Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Guatemala and Kuwait.
Derived failure modes
-
Unknown
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.